Home » products » Load Cells » BLH Nobel » KIS-2
Severity: Warning
Message: Undefined property: stdClass::$SchedaTecnica
Filename: views/dettaglio_prodotto.php
Line Number: 40
Load Cells
BLH Nobel
Severity: Warning
Message: Undefined property: stdClass::$SchedaTecnica
Filename: views/dettaglio_prodotto.php
Line Number: 59




| Rated load (R.L.): | 1, 2, 5, 10, 20, 30 kN |
| Combined error: | ±0.05% of R.O. |
| Repeatability: | ±0.01% of R.O. |
| Safe overload: | 200*, 150* for 30 kN % of R.L. |
| Ultimate overload: | 300*, 200 for 30 kN % of R.L. |
| Safe uplift: | 100% of R.L. |
| Uplift Ultimate: | 120% of R.L. |
| Safe sideload: | 100*, 50* for 30kN % of R.L. |
| Ultimate Sideload: | 200*, 100 for 30 kN % of R.L. |
| Recommended input: | 5 Vdc or Vac |
| Input voltage maximum: | 18 Vdc or Vac |
| Input resistance: | 350 ± 3 Ohms |
| Output resistance: | 350 ± 3 Ohms |
| Rated output (R.O.): | 2.040 mV/V ±0.25% |
| Zero balance: | ±5% of R.O. |
| Creep (30 minutes): | ±0.03% of R.L. |
| Temperature range: | -40 ÷ +100*** °C *** on request |
| Temp. effect (-10°C to +50°C) on output: | ±0.0033% of output/°C |
| Temp. effect (-10° to +50°C) on zero balance: | ±0.0014% of R.O./°C |
| Insulation resistance at 200V DC: | > 4 G Ohms |
| Material (load cell): | Stainless steel |
| Material (accessories): | Yellow chromate steel or stainless steel |
| Electrical connection: | 10 m shielded four conductor cable (50 kN) |
| Degree of protection: | IP67 |
| ATEX/FM/CSA | Certified versions for use in explosive atmospheres are available |
| ATEX: | II 1 GD EEx ia, IIC T4 Tamb = 40 °C EEx ia, IIC T6 Tamb = 60 °C |
| FM: | 3611 (Classes I, II, III; Div 1,2; Group A-G) |
| CSA: | C22.2 (Classes I, II, III; Div 1, 2; Group A-G) |
| * With reference to the loading point recommended |