Home » products » Load Cells » Bending beam » ALPHA
 
    Load Cells
Bending beam
 
    		



| Rated load (R.L.): | 25, 50, 100, 150, 250, 500 N | 
| Combined error: | < ±0.02% of R.O. | 
| Repeatability: | < ±0.01% of R.O. | 
| Creep (20 minutes): | ±0.05% of R.L. | 
| Safe overload: | 175% of R.L. | 
| Ultimate overload: | 300% of R.L. | 
| Recommended input: | 5÷12 Vdc/Vac | 
| Maximum excitation: | 20 Vdc/Vac | 
| Input resistance: | 350 ± 3.5 Ohm | 
| Output resistance: | 350 ± 3.5 Ohm | 
| Insulation resistance: | 2000 M Ohm | 
| Rated output (R.O.): | 3.0 mV/V | 
| Zero balance: | < ±2% of R.O. | 
| Temperature range: | -25 ÷ +80 °C | 
| Compensated temperature range: | -15 ÷ +65 °C | 
| Temperature effect on zero balance: | < ±0.0014% of R.O./°C | 
| Temperature effect on output: | < ±0.0014% on output/°C | 
| Max load deflection: | 0.254 mm capacity 25÷250 N; 0.432 mm capacity 500 N | 
| Degree of protection: | IP67 | 
| Material: | copper and tin-plated brass | 
| Electrical connection: | 4 conductors shielded cable; length 1.5 m | 
| Certifications: | OIML EN 45501: 1992 (25÷250 N), CSA C22.2, FM3611 | 